True RMS model BM257 ACV & ACA accuracies are specified from 5 % to 100 % of range or otherwise specified.  
		
		
			  
		
		
			Maximum Crest Factor <3:1 at full scale & <6:1 at half scale, and with frequency components fall within the meter specified frequency bandwidth for non-sinusoidal waveforms  
		
		
			  
		
		
			AC & AC+DC Voltage  
		
	 
 
	
		
			
				
					| 
							Function  
						 | 
							RANGE  
						 | 
							Accuracy  
						 | 
				
					| 
							50Hz ~ 60Hz  
						 | 
				
					| 
							mV  
						 | 
							60.00mV, 600.0mV  
						 | 
							0.5% + 3d  
						 | 
				
					| 
							V  
						 | 
							9.999V, 99.99V, 999.9V  
						 | 
				
					| 
							40Hz ~ 500Hz  
						 | 
				
					| 
							mV  
						 | 
							60.00mV, 600.0mV  
						 | 
							0.8% + 4d  
						 | 
				
					| 
							V  
						 | 
							9.999V, 99.99V  
						 | 
							1.0%+4d  
						 | 
				
					| 
							999.9V  
						 | 
							2.0%+4d  
						 | 
				
					| 
							500Hz ~ 1kHz  
						 | 
				
					| 
							mV  
						 | 
							60.00mV, 600.0mV  
						 | 
							2.0% + 3d  
						 | 
				
					| 
							V  
						 | 
							9.999V, 99.99V  
						 | 
							1.0%+4d  
						 | 
				
					| 
							999.9V  
						 | 
							2.0%+4d  
						 | 
				
					| 
							1kHz ~ 3kHz  
						 | 
				
					| 
							mV  
						 | 
							60.00mV, 600.0mV  
						 | 
							2%+3d  
						 | 
				
					| 
							V  
						 | 
							9.999V, 99.99V, 999.9V  
						 | 
							3.0%+4d  
						 | 
				
					| 
							3kHz ~ 20kHz  
						 | 
				
					| 
							mV  
						 | 
							60.00mV 1), 600.0mV 1)  
						 | 
							2%+3d  
						 | 
				
					| 
							V  
						 | 
							9.999V, 99.99V  
						 | 
							3dB  
						 | 
				
					| 
							999.9V  
						 | 
							Unspec’d  
						 | 
			
		
		
			1) Specificied from 30% to  of range.  
		
		
			CMRR:>60dB @ DC to 60Hz, Rs=1k W  
		
		
			Input Impedance: 10M W , 50pF nominal (80pF nominal for 600mV range)  
		
		
			Residual reading less than 5 digits with test leads shorted.  
		
		
			  
		
		
			AutoCheckTM (ACV)  
		
		
			
				
					| 
							RANGE 1)  
						 | 
							Accuracy  
						 | 
				
					| 
							50Hz ~ 60Hz  
						 | 
				
					| 
							9.999V, 99.99V, 999.9V  
						 | 
							1.0%+4d  
						 | 
			
		
	 
 
	
		Lo-Z ACV Threshold:> 3VAC (50/60Hz) nominal  
	
	
		Lo-Z ACV Input Impedance:  
	
	
		Initially approx. 3.0k W , 150pF nominal; Impedance increases abruptly within 
a fraction of a second as display voltage is above 50V (typical). Ended up impedances 
vs display voltages typically are:  
	
	
		18k W       @100V           125k W       @ 300V  
	
	
		320k W     @ 600V          460k W        @ 1000V  
	
	
		  
	
	
		dBm  
	
	
		At 600 W , -11.76dBm to 54.25dBm,  
	
	
		Accuracy: ± 0.25dB + 2d (@40Hz -- 20kHz)  
	
	
		Input Impedance: 10M W , 50pF nominal  
	
	
		Selectable reference impedance of  4, 8, 16, 32, 50, 75, 93, 110, 125, 135, 150, 200, 
250, 300, 500, 600, 800, 900, 1000, 1200 W  
	
	
		  
	
	
		DC Voltage  
	
	
		
			
				| 
						Function  
					 | 
						RANGE  
					 | 
						Accuracy  
					 | 
			
				| 
						mV  
					 | 
						60.00mV  
					 | 
						0.12%+2d  
					 | 
			
				| 
						600.0mV  
					 | 
						0.06%+2d  
					 | 
			
				| 
						V  
					 | 
						9.999V, 99.99V, 999.9V  
					 | 
						0.08%+2d  
					 | 
		
	
	
		NMRR:        >60dB    @ 50/60Hz  
	
	
		CMRR:       >110dB @ DC, 50/60Hz, Rs=1k W  
	
	
		Input Impedance: 10M W , 50pF nominal (80pF nominal for 600mV range)  
	
	
		  
	
	
		AutoCheckTM (DCV)  
	
	
		
			
				| 
						RANGE 
					 | 
						Accuracy  
					 | 
			
				| 
						9.999V, 99.99V, 999.9V  
					 | 
						0.5%+3d  
					 | 
		
	
	
		Lo-Z DCV Threshold:  
	
	
		 > +1.5VDC or < -1.0VDC nominal  
	
	
		Lo-Z DCV Input Impedance:  
	
	
		Initially approx. 3.0k W , 165pF nominal; Impedance increases abruptly within a fraction 
of a second as display voltage is above 50V (typical). Ended up impedances vs display 
voltages typically are:  
	
	
		18k W       @100V          125k W         @ 300V  
	
	
		320k W     @ 600V         500k W         @ 1000V  
	
	
		  
	
	
		Ohms  
	
	
		
			
				| 
						RANGE 
					 | 
						Accuracy  
					 | 
			
				| 
						600.0 W , 6.000k W , 60.00k W , 600.0k W  
					 | 
						0.1%+3d  
					 | 
			
				| 
						6.000M W  
					 | 
						0.4%+3d  
					 | 
			
				| 
						60.00M W  
					 | 
						1.5%+5d  
					 | 
			
				| 
						99.99nS  
					 | 
						0.8%+10d  
					 | 
		
	
	
		Open Circuit Voltage: < 1.2VDC (< 1.0VDC for 60M W range)  
	
	
		  
	
	
		AutoCheckTM (Ohms)  
	
	
		
			
				| 
						RANGE 
					 | 
						Accuracy  
					 | 
			
				| 
						600.0 W , 6.000k W , 60.00k W , 600.0k W  
					 | 
						0.5%+4d  
					 | 
			
				| 
						6.000M W  
					 | 
						0.8%+3d  
					 | 
			
				| 
						60.00M W  
					 | 
						2.0%+5d  
					 | 
		
	
	
		Open Circuit Voltage: < 1.2VDC (< 1.0VDC for 60M W range)  
	
	
		  
	
	
		Audible Continuity Tester  
	
	
		Audible threshold: between 20 W and 300 W ; Response time < 100 m s  
	
	
		  
	
	
		Diode Tester  
	
	
		
			
				| 
						RANGE 
					 | 
						Accuracy  
					 | 
			
				| 
						2.000V  
					 | 
						1.0%+1d  
					 | 
		
	
	
		Test Current (Typically): 0.4mA  
	
	
		Open Circuit Voltage: < 3.5 VDC  
	
	
		  
	
	
		Temperature  
	
	
		
			
				| 
						RANGE 
					 | 
						Accuracy  
					 | 
			
				| 
						-50oC to 1000oC  
					 | 
						0.3%+2oC  
					 | 
			
				| 
						-58oF to 1832oF  
					 | 
						0.3%+5oF  
					 | 
		
	
	
		Type-K thermocouple range & accuracy not included  
	
	
		  
	
	
		Capacitance  
	
	
		
			
				| 
						RANGE 
					 | 
						Accuracy 1)  
					 | 
			
				| 
						60.00nF, 600.0nF  
					 | 
						0.8% + 3d  
					 | 
			
				| 
						6.000 m F  
					 | 
						1.0% + 3d  
					 | 
			
				| 
						60.00 m F  
					 | 
						2.0% + 3d  
					 | 
			
				| 
						600.0 m F 2)  
					 | 
						3.5% + 5d  
					 | 
			
				| 
						6.000mF 2)  
					 | 
						5.0% + 5d  
					 | 
			
				| 
						25.00mF 2)  
					 | 
						6.5% + 5d  
					 | 
		
	
	
		1) Accuracies with film capacitor or better  
	
	
		2) In manual-ranging mode, measurements not specified below 50.0 mF, 0.54mF and 
5.4mF for 600.0 mF, 6.000mF and 25.00mF ranges respectively  
	
	
		  
	
	
		AC & AC+DC Current  
	
	
		
			
				| 
						RANGE  
					 | 
						Accuracy  
					 | 
						Burden voltage  
					 | 
			
				| 
						BM82x 2)  
					 | 
						BM52y 3)  
					 | 
			
				| 
						50Hz ~ 60Hz  
					 | 
			
				| 
						600.0 m A, 6000 mA  
					 | 
						0.6%+3d  
					 | 
						0.08mV/ m A  
					 | 
						0.08mV/ m A  
					 | 
			
				| 
						60.00mA  
					 | 
						1.5mV/mA  
					 | 
						2.1mV/mA  
					 | 
			
				| 
						600.0mA  
					 | 
						1.0%+3d  
					 | 
			
				| 
						6.000A , 10.00A 1)  
					 | 
						0.8%+6d  
					 | 
						0.04V/A  
					 | 
						0.02V/A  
					 | 
			
				| 
						40Hz ~ 1kHz  
					 | 
			
				| 
						600.0 m A, 6000 mA  
					 | 
						0.8%+4d  
					 | 
						0.08mV/ m A  
					 | 
						0.08mV/ m A  
					 | 
			
				| 
						60.00mA  
					 | 
						1.5mV/mA  
					 | 
						2.1mV/mA  
					 | 
			
				| 
						600.0mA  
					 | 
						1.0%+4d  
					 | 
			
				| 
						6.000A , 10.00A 1)  
					 | 
						0.8%+6d  
					 | 
						0.04V/A  
					 | 
						0.02V/A  
					 | 
		
	
	
		1) 10A continuous, > 10A to 15A (to 20A for BM521 & BM525) for 30 second max with 5 minutes cool down interval  
	
	
		2) x=1, 2, 7 & 9; Model 821, 822, 827 & 829  
	
	
		3) y=1 & 5; Model 521 & 525  
	
	
		  
	
	
		DC Current  
	
	
		
			
				| 
						RANGE 
					 | 
						Accuracy  
					 | 
						Burden voltage  
					 | 
			
				| 
						BM82x 2)  
					 | 
						BM52y 3)  
					 | 
			
				| 
						600.0 m A, 6000 mA  
					 | 
						0.2%+4d  
					 | 
						0.08mV/ m A  
					 | 
						0.08mV/ m A  
					 | 
			
				| 
						60.00mA, 600.0mA  
					 | 
						1.5mV/mA  
					 | 
						2.1mV/mA  
					 | 
			
				| 
						6.000A , 10.00A 1)  
					 | 
						0.04V/A  
					 | 
						0.02V/A  
					 | 
		
	
	
		1) 10A continuous, > 10A to 15A (to 20A for BM521 & BM525) for 30 second max with 5 minutes cool down interval  
	
	
		2) x=1, 2, 7 & 9; Model 821, 822, 827 & 829  
	
	
		3) y=1 & 5; Model 521 & 525  
	
	
		  
	
	
		Line Level Frequency (Hz)  
	
	
		
			
				| 
						Function Range  
					 | 
						Frequency  
					 | 
						Sensitivity (Sine RMS)  
					 | 
			
				| 
						AC 60.00mV  
					 | 
						15.00 ~ 50.00kHz  
					 | 
						40mV  
					 | 
			
				| 
						AC 600.0mV  
					 | 
						60mV  
					 | 
			
				| 
						AC 9.999V  
					 | 
						15.00 ~ 10.00kHz  
					 | 
						2.5V  
					 | 
			
				| 
						AC 99.99V  
					 | 
						25V  
					 | 
			
				| 
						AC 999.9V  
					 | 
						100V  
					 | 
			
				| 
						AC 600.0 mA  
					 | 
						  
					 
						  
					 
						15.00 ~ 3.000kHz  
					 | 
						45uA  
					 | 
			
				| 
						AC 6000 mA  
					 | 
						600uA  
					 | 
			
				| 
						AC 60.00mA  
					 | 
						40mA  
					 | 
			
				| 
						AC 600.0mA  
					 | 
						60mA  
					 | 
			
				| 
						AC 6.000A  
					 | 
						4A  
					 | 
			
				| 
						AC 10.00A  
					 | 
						6A  
					 | 
		
	
	
		Accuracy: 0.04%+4d  
	
	
		Crest mode (Instantaneous Peak Hold)  
	
	
		Accuracy: Specified accuracy adds 250 digits for changes > 1.0 ms in duration  
	
	
		  
	
	
		Record mode  
	
	
		Accuracy: Specified accuracy adds 10 digits for changes > 100 ms in duration  
	
	
		  
	
	
		Logic Level Frequency ( Hz) &  Duty Cycle (D%)  
	
	
		
			
				| 
						@ DCmV Function  
					 | 
						Range  
					 | 
						Accuracy 1)  
					 | 
			
				| 
						Frequency  
					 | 
						5.00Hz ~ 1.000MHz  
					 | 
						0.004%+4d  
					 | 
			
				| 
						Duty Cycle  
					 | 
						0.00% ~ 100.0%  
					 | 
						3d/kHz+2d 2)  
					 | 
		
	
	
		1) Sensitivity: 2.5Vp (Square wave) for 3V & 5V Logic Family  
	
	
		2) Specified Frequency: 5Hz ~ 10kHz 
	
	
		  
	
	
		Non-Contact EF-Detection  
	
	
		
			
				| 
						Typical Voltage  
					 | 
						Bar Graph Indication  
					 | 
			
				| 
						20V (tolerance:10V~36V)  
					 | 
						-  
					 | 
			
				| 
						55V (tolerance: 23V ~ 83V)  
					 | 
						--  
					 | 
			
				| 
						1  10V (tolerance: 59V ~ 165V)  
					 | 
						---  
					 | 
			
				| 
						220V (tolerance: 124V ~ 330V)  
					 | 
						----  
					 | 
			
				| 
						440V (tolerance: > 250V)  
					 | 
						-----  
					 | 
		
	
	
		Indication: Bar graph segments & audible beep tones proportional to the field strength  
	
	
		Detection Frequency: 50/60Hz  
	
	
		Detection Antenna: Top end of the meter  
	
	
		Probe-Contact EF-Detection:  
	
	
		For more precise indication of live wires, such as distinguishing between live and 
ground connections, use the Red (+) test probe for direct contact measurements.